HIGH SPEED HANDPIECE MAINTENANCE GUIDELINES

3. Re-connect handpiece to unit tubing, place an autoclave bag over the handpiece & run 30 seconds to eliminate excess oil from the turbine. (This procedure is primarily to clean the handpiece not lubricate it).
4. Remove from unit tubing, and autoclave.
5. After autoclaving, allow to cool gradually, about 30 minutes, to room temperature.
6. With bur in handpiece place lubricant (2 or 3 drops) into air drive port, - at the same time, rotate the bur between your fingers forward/backward 1/4 turns for 8 to 10 seconds.
7. Connect handpiece to unit tubing & run for 30 seconds to eliminate excess oil from the turbine (this procedure is to lubricate the handpiece).

LUBRICATING CONTRA ANGLES AND HEADS

3. Place a lubricant (2 or 3 drops) into all parts noted by the arrow.
4. Reassemble, the Contra Angle and place it on a Slow speed motor and run for 30 seconds (this is for cleaning purposes).
5. Autoclave and let cool.
6. Repeat steps 1 thru 4 (this is for lubricating purposes).